HoMedics BPW-370BT Instruction for Use Page 17

  • Download
  • Add to my manuals
  • Print
  • Page
    / 39
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 16
32
Note: If the unit still does not work, contact HoMedics Consumer Relations. Under no
circumstance should you disassemble or attempt to repair the unit by yourself.
Contact information for HoMedics Consumer Relations Department can be found
on the warranty page.
TROUBLESHOOTING
If any abnormality arises during use, please check the following points:
SYMPTOMS POSSIBLE CAUSES CORRECTION
Unit does not turn
on when the START/
STOP button is
pushed.
Batteries have run down.
Replace them with two new “AAA”
alkaline batteries.
Battery polarities
have been positioned
incorrectly.
Re-insert the batteries in the correct
positions.
EE measurement
error symbol shown
on display or the
blood pressure
value is displayed
excessively low (or
high).
Cuff has been placed
incorrectly.
Wrap the cuff properly so that it is
positioned correctly. Measure again.
Did you talk or move
during the measurement?
Keep wrist steady during
measurement. Measure again.
Shaking of the wrist with
the cuff on.
E1 error symbol
shown on display.
Air circuit abnormality. Measure again.
E2 error symbol
shown on display.
Inflation pressure
exceeding 300mmHg.
Switch the unit off, then measure
again.
E3 error symbol
shown on display.
Error determining
measurement data
Measure again.
EP error symbol
shown on display.
System error. Measure again.
Monitor cannot
communicate with
Bluetooth-enabled
device
Bluetooth-enabled device
does not pair with BPM
or is paired with a wrong
device.
Re-pair the BPM and Bluetooth-
enabled device with each other.
Bluetooth function is not
turned ON.
Make sure the device is in standby
mode.
The distance between
BPM and Bluetooth-
enabled device is out of
transmitting range.
Make sure the devices are at an
acceptable distance (32 feet/10
meters) with each other.
Using an incompatible
Bluetooth-enabled device.
Refer to Page 21 “Bluetooth
compatibility” and Page 30
“RF Specification”
Using non-Bluetooth-
enabled device.
Unexpected loss of
electrical/mechanical
integrity.
Re-insert the batteries and try again.
Return the device to
your local distributor
or importer.
31
• Guidance and manufacturer’s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments listed below, and
should only be used in such environments:
Immunity
test
IEC
60601
test level
Compliance
level
Electromagnetic environment – guidance
Conducted
RF
IEC 61000-
4-6
Radiated
RF
IEC 61000-
4-3
3 Vrms
150 kHz
to 80
MHz
3 V/m
80 MHz to
2.5 GHz
Not
Applicable
3 V/m
Portable and mobile RF communications equip-
ment should be used no closer to any part of the
device, including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
d = 1.2
80 MHz to 800 MHz
d = 2 .3 800 MHz to 2.5 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separa-
tion distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey, a
should be less than the compliance level in each
frequency range.b
Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be con-
sidered. If the measured field strength in the location in which the device is used exceeds
the applicable RF compliance level above, the device should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary,
such as reorienting or relocating the device.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Page view 16
1 2 ... 12 13 14 15 16 17 18 19 20 21 22 ... 38 39

Comments to this Manuals

No comments